1. Department of Oncology and Radiotherapy, Xiang’an Hospital Affiliated to Xiamen University, Xiamen 361100, China
2. Department of Medical Records, Xiang’an Hospital Affiliated to Xiamen University, Xiamen 361100, China
| Abstract: | Background: Our hospital has been using Varian VitalBeam linear accelerator since December 2020. Daily quality control of accelerator beam stability in precision radiotherapy is very important. In this paper, we compare and analyze the beam stability of the accelerator with two monitoring methods. Objective: The long-term stability analyze and study of VitalBeam linear accelerator by using machine Machine Performance Check (MPC) and Morning Check. Methods: The energies of 6MV and 10MV x-ray most commonly used by Varian Linear accelerator (VitalBeam) for patient treatment were analyzed and compared,Firstly, the absolute dose of 6MV and 10MV X-ray energies of VitalBeam was calibrated. Collected and analyzed the morning checking data of VitalBeam from November 2021 to May 2022, including the CAX, SYM and FLAT offset data; the MPC daily data was collected and analyzed at the same time, which include beam output change, uniformity change and beam center shift data. The beam output changes were statistically analyzed, and the stability of other data was analyzed. Results: The Morning Check and MPC can accurately detect the change of accelerator beam performance, and the measured values are stable and within the tolerance range,The comparison of 6MV and 10MV beam outputs between the morning detector and MPC was statistically significant (P<0.05). Conclusion: Both MPC and Morning Check can detecte the requirements of accelerator quality assurement, the beam output monitoring morning detector has advantages, but MPC has faster execution efficiency than morning detector instrument, and is more suitable for units with a large number of patients treated daily. |
| Keywords: | Morning Detector; MPC; Linear Accelerator; Beam Stability; Daily Inspection |
| DOI: | 10.57237/j.se.2023.02.001 |
| 1. | 厦门市医疗卫生指导项目 (3502Z20209111) |
| [1] | 王哲, 程金生. 晨检仪在直线加速器日常质量控制应用中的性能研究 [J]. 中国医学装备, 2018, 15 (4): 43-47. |
| [2] | 习海燕.医用直线加速器TrueBeam剂量输出长期稳定性[J].医疗装备, 2019, 32 (17): 28-29. |
| [3] | KUTCHER G J, COIA L, GILLIN M, et al. Comprehensive QA for radiation oncology: report of AAPM radiation therapy committee task group 40 [J]. Med Phys, 1994, 21 (4): 581-618. |
| [4] | KLEIN E E, HANLEY J, BAYOUTH J, et al. Task Group 142 report: quality assurance of medical accelerators [J]. Med Phys, 2009, 36 (9): 4197-4212. |
| [5] | 赖友群.基于MPC的VitalBeam直线加速器束流一致性与准直系统性能检测 [J]. 中国医疗器械信息, 2020, 26 (18): 177-180. |
| [6] | 杨玉敏.基于MPC分析VitalBeam加速器的稳定性[J].中国医学工程, 2022, 30 (9): 1-4. |
| [7] | 雷伟杰, 王斌, 韩理想. MPC在Vital Beam加速器性能检测中的应用 [J]. 中国医疗设备, 2020, 35 (11): 73-76. |
| [8] | 蒋大振, 王骁踊, 张俊.运用PTW QUICKCHECK webline晨检仪分析医用加速器输出稳定性 [J]. 中国医学物理学杂志, 2019, 36 (5): 540-545. |
| [9] | 易海云, 夏兵, 姚雷,等.照射野等中心的偏移对 VMAT 计划剂量准确性的影响 [J]. 中国医学物理学杂志, 2017, 34 (8): 794-798. |
| [10] | 柳先锋,刘翔宇,何亚男,等. Varian On-Board Imager (OBI)常规测量方法及质量保证 [J]. 中国医学物理学杂志, 2011, 28 (1): 2344-2347. |
| [11] | 张基永, 吴丽丽, 林珠, 等.基于MIMI模体对Truebeam 加速器影像系统的质量保证 [J]. 中国医学影像学杂志, 2015, 23 (7): 557-560. |
| [12] | 叶淑敏, 滕建建, 石锦平等. MLC 叶片系统误差对鼻咽癌 VMAT和 IMRT 计划剂量影响的比较 [J]. 中国医学物理学杂志, 2019, 36 (10): 1139-1144. |
| [13] | 马蕾杰, 宋迎新, 雷宏昌等. XHA600E型国产数字化加速器性能检测与对比评价 [J]. 中国医疗设备, 2018, 33 (6): 88-91. |
| [14] | 姜斐, 于浪, 孙显松等. 基于机器性能检查的TrueBeam直线加速器射线稳定性研究 [J]. 中国医学装备, 2022, 19 (5): 13-17. |
| [15] | 吴丽丽, 林浩, 陈泓, 等.医用加速器TrueBeam剂量输出长期稳定性研究 [J]. 中国医学装备, 2014, 11 (6): 34-37. |